Collector Supply
Peak Voltage (Range) | 16 V | 80 V | 400 V | 2,000 V |
---|---|---|---|---|
Polarity Modes | AC, ± DC, ± Leakage, ± Rectified Sine | |||
Available Current | ||||
Peak Current | 10 A | 2 A | 0.4 A | 0.05 A |
Peak Current (PULSE mode) | 20 A | 4 A | 0.8 A | 0.1 A |
Available Resistance | ||||
Minimum Series Resistance | 0.26 Ω | 6.4 Ω | 160 Ω | 20 kΩ |
Maximum Series Resistance | 800 Ω | 20 kΩ | 500 kΩ | 12.5 MΩ |
Available Peak Power
16, 80 and 400 V Range - 0.08, 0.4, 2, 10, 50, 220 W.
2,000 V Range - 0.08, 0.4, 2, 10, 50 W.
Variable Collector Supply Settings As a % of Maximum Peak Voltage - 0 to 100% in 0.1% increments.
Looping Compensation (typical) Range - at least 100 pF.
Step Generator
Accuracy
Incremental - 1.5%.
Absolute - Less than 1.5% of total output + 3% x STEP AMPLITUDE setting + 1 mV or 1 nA.
Offset Control Range - Variable from -10 to +10 times STEP AMPLITUDE.
Resolution - STEP AMPLITUDE setting x 1%.
Current Mode
Amplitude Range - 50 nA to 200 mA in a 1-2-5 sequence of 21 steps.
Maximum Current - 20x STEP AMPLITUDE setting, 10x STEP AMPLITUDE setting when control is set to 200 mA.
Maximum Voltage - At least 10 V.
Maximum Opposing Offset Current - 10x STEP AMPLITUDE.
Maximum Opposing Volts - Less than 15 V.
Ripple Plus Noise - Less than 0.5% x STEP AMPLITUDE + 10 nA.
Voltage Mode
Amplitude Switch Range - 50 mV to 2 V, in a 1-2-5 sequence of 6 steps.
Maximum Voltage - 20x STEP AMPLITUDE.
Maximum Current - At least 500 mA at 10 V or less, at least 200 mA at 15 V, at least 10 mA at 40 V.
Short Circuit Current Limiting - 500 mA, +50%, -20%.
Maximum Opposing Offset Current - 10x STEP AMPLITUDE.
Maximum Opposing Current - Less than 20 mA.
Ripple Plus Noise - Less than 0.5% x STEP AMPLITUDE + 10 mV.
Step Rates - 2x line frequency (1 x line frequency in AC collector supply mode). Steps occur at 0 collector voltage.
Pulsed Steps - 80 µs or 300 µs wide, ±10%.
Steps and Offset Polarity - Corresponds to Collector Supply Polarity when STEP GENERATOR POLARITY INVERT is disabled. Opposite to Collector Supply Polarity when STEP GENERATOR POLARITY INVERT is selected or CONFIGURATION switch is set to BASE COMMON. BASE COMMON configuration disables STEP GENERATOR POLARITY INVERT.